Pocket-DSO
DS-1000 series Specifications
|
Acquisition |
|
|
Mode |
Real-time
sampling, Equivalent sampling, Roll mode, Average, Envelope |
|
Input |
|
|
Input
Coupling |
AC,
DC, GND |
|
Input
Impedance |
1MΩ±1%
// 21pF±5% |
|
|
42Vpk
(DC + AC peak) |
|
Vertical |
|
|
Channel |
2
(DS-1202 allows to stack up to max. 6 channels.) |
|
Vertical
Resolution |
9
bits /channel @ 5mV/DIV- 10V/DIV (8 bits @ 2mV/DIV) |
|
Sensitivity |
2mV/DIV
to 10V/DIV (as 1-2-5 step) |
|
Bandwidth |
DS-1102, DS-1202: DC to 200MHz |
|
BW
Limit |
Approx.
20MHz |
|
Range |
8
divisions |
|
Offset
Level |
±4
divisions |
|
Offset
Increments |
0.1
division |
|
DC
accuracy |
±3% |
|
Time
Base |
|
|
Sampling
Rate |
DS-1102, DS-1202: Real-time
sampling: 200MS/s @ 1Ch, 100MS/s @2Ch (Single Shot) |
|
Time
Base Range |
5ns/DIV
to 10s/DIV (as 1-2- 5 step) |
|
Accuracy |
100ppm |
|
Range |
10
Divisions |
|
Delay Trigger |
320
Divisions (DS-1002,DS-1102) |
|
Time
Resolution |
200ps |
|
Record Length |
DS-1002, DS-1102: 2k points to
64k points |
|
Trigger |
|
|
Type |
Rising,
Falling, Delay-Trigger, TV-Trigger (DS-1002
no TV Trigger) |
|
Mode |
Auto,
Normal and Single (with
RUN/STOP hardware button on the DSO device) |
|
Source |
CH1,
CH2, Ext-Trig |
|
Coupling |
DC |
|
Sensitivity |
DC
to 25MHz = 0.35div or 3.5mv; |
|
Trigger
Level |
±4
divisions |
|
Trigger
Increments |
0.1
division |
|
Measurement
and Processing |
|
|
Special
function |
Autoset,
Monitor from Internet (TCP/IP) |
|
Measurement |
Vp-p,
Vmax, Vmin, Vamp, Vmean, Vrms, Period, Frequency, ±Duty, ±Pulse width |
|
Cursor |
Time
difference, Voltage difference |
|
Math |
Add,
Sub, Multiplication, Division |
|
FFT |
Rectangular
(Vertical scale: dbV RMS, Linear RMS) |
|
Export
Data |
WORD,
EXCEL, CSV, TEXT, HTML, Clipboard, Hardcopy, Preview |
|
Trigger
Input/Output |
|
|
EXT-TRIG
Input Limitation |
TTL
Level |
|
EXT-TRIG
Input Acknowledge Level |
1.6V
to 5V, rising/falling edge |
|
EXT-TRIG Input Acknowledge Time |
“>10ns”
and “>0.1 TIME/DIV” |
|
TRIG-OUT |
3.3v pulse, 20ns delay after trigger occurring |
|
Compensation
Output |
|
|
Level |
Approx.
3.3V |
|
Frequency |
1kHz
±0.5% |
|
Environment |
|
|
Operation |
0℃
to +50℃ |
|
Storage |
-10℃
to +60℃ |
|
Physical |
|
|
Interface |
USB2.0
(USB1.1 compatible) |
|
Power |
USB
bus power |
|
Dimension |
135/80/26
mm3 |
|
Weight |
230
g |
|
Accessories |
|
|
Probes |
DS-1102, DS-1202: 250MHz probe
(1x/10x) x2 |
|
Others |
Installation
CD, USB2.0 cable, User manual, Soft case |
DS-1202
Spec.
Limitations in Stack Mode
|
Acquisition |
|
|
Mode |
Equivalent sampling and Roll mode only available in Master-DSO |
|
Time
Base |
|
|
Sampling
Rate |
Real-time
sampling: the same as individual unit |
|
Time
Base Range |
5ns/DIV
to 100ms/DIV (as 1-2- 5 step) for Master and Slave DSOs;
|
|
Trigger |
|
|
Source |
CH1,
CH2, Ext-Trig in Master-DSO (NA for Slave-DSO) |
|
Jitter |
+/-
200ps in Master-DSO; +/-10ns in Slave-DSO |
DS-1000
System Requirement
|
1. |
INTEL Pentium 1.2G CPU
or compatible PC (2G recommended) |
|
2. |
One USB port (USB2.0
recommended) |
|
3. |
Windows
98SE/ME/2000/XP/2003
operating system |
|
4. |
128M bytes memory
(512M recommended) |
|
5. |
Disk space more than
10M bytes |
|
6. |
One CD-ROM drive for
installation |
|
7. |
800x600 VGA or higher
(1280x1024 recommended) |
|
8. |
101-Keys keyboard |
|
9. |
Mouse (with wheel
recommended) |
Packing List of
DS-1000 series
| Item | DS-1002/DS-1102 | DS-1202 | |
| 1. | Pocket DSO device | 1 | |
| 2. | 250MHz probes (with 1x/10x switch) | 2* | |
| 3. | Probe accessories pack | 2 | |
| 4. | Stack function cable | 0 | 1 |
| 5. | USB2.0 1.8m cable | 1 | |
| 6. | Installation CD | 1 | |
| 7. | User manual | 1 | |
| 8. | DSO soft case | 1 | |
| 1. | 2. | 3. | 4. | 5. | 6. |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
*Note:DS-1002 provides 2 pieces of 100MHz probes (1x/10x).
Specification and Equipment PDF file download
DS-1000
Driver & Manual download (point
on the picture click mouse right buttom to save)
![]() |